Some Technical Definitions
ABSOLUTE PRESSURE TRANSDUCER:
A pressure transducer or pressure sensor that has an internal
reference chamber sealed at or close to 0 psia (full vacuum) and normally
provides increasing output voltage for increases in pressure.
ACCURACY:
The combined error of nonlinearity, repeatability, and hysteresis
expressed as a percent of full scale output.
AXIAL LOAD:
A load applied along or parallel to and concentric with the
primary axis.
BRIDGE:
A Wheatstone bridge configuration utilizing four active strain
gages.
BRIDGE RESISTANCE:
The nominal value of the individual legs that make up a complete
Wheatstone Bridge.
CALIBRATION:
The comparison of transducer voltage outputs against the outputs
of a reference standard.
DAMPING:
The reduction of response at the resonant frequency through the
use of a damping media such as oil. Usually specified as the ratio of critical
damping.
DEAD VOLUME:
The volume inside the pressure port of a transducer at room
temperature and barometric pressure.
DEFLECTION:
The change in length along the primary axis or distance a
diaphragm moves at the center between no-load and rated load conditions.
DIAPHRAGM:
The sensing membrane which is deformed when pressure is applied.
EXCITATION, ELECTRICAL:
The voltage or current applied to the input terminals of the
transducer.
FLUSH DIAPHRAGM:
Sensing element is located on the very tip of the transducer (NO
pressure port).
FREQUENCY RESPONSE:
The range of frequencies over which the transducer voltage output
will follow the sinusoidally varying mechanical input within specified limits.
FULL SCALE:
See Rated Capacity.
FULL SCALE OUTPUT:
The algebraic difference between the minimum output (normally
zero) and the rated capacity.
GAGE PRESSURE:
The pressure above (or below) atmospheric. Represents positive
difference between measured pressure and existing atmospheric pressure. Can be
converted to absolute by adding actual atmospheric pressure value.
GAGE PRESSURE TRANSDUCER:
A transducer which measures pressure relative to the atmospheric
pressure.
HYSTERESIS:
The maximum difference between output readings for the same
measured point, one point obtained while increasing from zero and the other
while decreasing from full scale. The points are taken on the same continuous
cycle. The deviation is expressed as a percent of full scale.
INPUT IMPEDANCE:
The resistance measured across the excitation terminals of a
transducer at room temperature, with no load applied, and with the output
terminals open-circuited.
INSULATION (ISOLATION) RESISTANCE:
The DC resistance expressed in ohms measured between any
electrical connector pin or lead wire and the transducer body or case. Normally
measured at 50 VDC.
LINEARITY:
The maximum deviation of the calibration curve from a straight
line between zero and full scale, expressed as a percent of full scale output
and measured on increasing measured only.
LINE PRESSURE:
The maximum pressure in the pressure vessel or pipe for
differential pressure measurement.
LOAD:
The weight, torque, or force applied to the transducer.
LOAD BUTTONS:
The spherical like shape of the top surface of a load cell where
the load is applied.
MEASURED MEDIA:
The physical quantity, property, or condition which is measured.
(eg: pressure, load, weight, acceleration)
MOUNTED RESONANT FREQUENCY:
The frequency at which the internal spring/mass system of an
accelerometer resonates, producing a 90º phase shift in output signal vs.
applied acceleration.
OUTPUT:
The electrical signal measured at the output terminals which is
produced by an applied input to a transducer.
OUTPUT IMPEDANCE:
The resistance as measured on the output terminals of a transducer
at standard temperature, with no measured applied, and with the excitation terminals
open-circuited.
OVERRANGE, SAFE:
The maximum pressure or load which may be applied to the
transducer without causing a permanent change in the performance
specifications.
PHASE SHIFT:
The phase angle between the output signal and the applied acceleration.
PRIMARY AXIS:
The axis along which the transducer is designed to be loaded;
normally its geometric centerline.
PSI:
Pounds per square inch.
PSIA:
Pounds per square inch absolute.
PSID:
Pounds per square inch differential.
PSIG:
Pounds per square inch gage.
PULL PLATE:
Load cell attachment which allows tension or compression force to
be directed at the center line of a load cell through a threaded center hole.
RANGE:
The measured values, over which a transducer is intended to
measure, specified by their upper and lower limits.
RATED CAPACITY:
The maximum measurand that a transducer is designed to measure
within its specification.
REPEATABILITY:
The ability of a transducer to reproduce output readings when the
same measured value is applied to it consecutively, under the same conditions,
and in the same direction. Repeatability is expressed as the maximum difference
between output readings as a percent of full scale.
RESOLUTION:
The smallest change in mechanical input which produces a
detectable change in the output signal.
SENSING ELEMENT:
The part of the transducer which reacts directly in response to
the measurand.
SENSITIVITY:
The ratio of change in transducer output to a change in the value
of the measurand.
SHUNT CAL (R-CAL):
The change in electrical output caused by placing a fixed resistor
between the appropriate transducer terminals. Used "in the field" for
quick calibration.
SPAN:
The algebraic difference between the limits of the range from zero
to full scale.
SPECIFICATIONS:
The group of error limits within which each device will operate.
STRAIN GAGE:
A measuring element for converting force, pressure, tension, etc.,
into an electrical signal.
TEMPERATURE, COMPENSATED:
The range of temperature over which a transducer can operate up to
full scale and still meet all specifications.
TEMPERATURE COMPENSATION:
The utilization of supplementary devices, materials, or components
within the bridge to minimize sources of error caused by changing temperature.
TEMPERATURE, OPERATING:
The range of temperature over which a transducer may be safely
operated up to full scale without causing failure, but specifications may not
be met.
TEMPERATURE EFFECT ON SPAN:
The change in rated output due to a change in ambient temperature.
Usually expressed as ± a percentage change in rated output per degree F change
in ambient temperature, over the compensated temperature range.
TEMPERATURE EFFECT ON ZERO:
The change in zero balance due to a change in ambient temperature.
Usually expressed as ± a percentage change in rated output per degree F change
in ambient temperature over the compensated temperature range.
TRANSDUCER:
A device (or medium) that converts energy from one form to
another. The term is generally applied to devices that take physical phenomenon
(pressure, temperature, humidity, flow, etc.) and convert it to an electrical
signal.
TRANSMITTER:
A transducer that has a 4-20 mA two wire output.
TRANSVERSE SENSITIVITY:
Signal output as a result of acceleration perpendicular to the
sensitive axis. Specified as a percentage of sensitive axis output for
equivalent right angle acceleration or as a decimal fraction.
VIBRATION ERROR:
The maximum change in output of a transducer when a specific
amplitude and range of frequencies are applied to a specific axis at room
temperature.
WET/DRY DIFFERENTIAL:
A differential pressure transducer or transmitter that uses a
metal diaphragm at the wet port where fluids can be applied, and no diaphragm
at the dry port. The dry port exposes the internal circuitry to the medium, so
only clean dry gas can be applied to this port.
WET/WET DIFFERENTIAL:
A differential pressure transducer or transmitter that has a metal
diaphragm in each pressure port to permit fluid into both ports.
WETTED PARTS:
The diaphragm and pressure port material that comes in direct
contact with the medium (gas, liquid).
ZERO ADJUSTMENTS:
Used when `setting up' a transducer to adjust the output signal to
zero when zero load/pressure is applied.
ZERO BALANCE:
The output signal of the transducer with rated excitation and with
no-load applied, usually expressed as a percent of rated output.
ZERO RETURN:
The difference in zero balance measured immediately before rated
load application of specified duration and measured after removal of the load,
and when the output has stabilized.